Technical data (Cattani)
| Rated voltage | 230 V |
| Rated frequency | 50/60 Hz |
| Supply phases | 1 |
| Maximum flow | 1000 l/min |
| Default vacuum level setpoint step 1 minimum level | 70 mbar |
| Default vacuum level setpoint step 2 medium level | 100 mbar |
| Default vacuum level setpoint step 3 maximum level | 210 mbar |
| Sound pressure level with canalized air | 59 dB(A) — Sound pressure level with canalized air 59 dB (A) |
| Sound pressure level measurement method | Sound pressure level detected according to ISO 3746-1979 (E). Parameters: r = 1.5, background noise 34 dB(A), Bruel & Kjaer type 2232 instrument. — Sound pressure level detected according to ISO 3746-1979 (E). Parameters: r= 1.5 - background noise: 34 dB(A) - Bruel & Kjaer type 2232 instrument. |
| Sound pressure level test method | Sound pressure level tested according to the standard ISO 3746-1979 (E). Parameters: r= 1.5 – background noise: 34 dB (A) |
| Sound pressure with box for indoors | 59 dB(A) |
| Width | 420 mm |
| Depth | 350 mm |
| Dimensions H | 470 mm |
| Net Weight | 23 kg |
| Gross Weight | 31 kg |
| Protection against humidity | Common |
| Transport storage max humidity | Packages cannot be exposed to water and spraying and cannot withstand humidity above 70%. % |
| Default blower current limit | 3,7 A |
| Blower target frequency | 60 Hz |
| Default blower target frequency step 2 medium level | 120 Hz |
| Default blower target frequency step 3 maximum level | 145 Hz |
| Default centrifuge separator current limit | 3 A |
| Default centrifuge separator frequency | 75 Hz |
| Diagnostic test inlet closed centrifuge separator current | 0,8 - 1,2 A |
| Diagnostic test inlet closed frequency reached | 80 - 95 Hz |
| Diagnostic test inlet closed head reached | 190 - 210 mbar |
| Diagnostic test inlet closed total current | 2,2 - 3 A |
| Diagnostic test inlet open centrifuge separator current | 0,8 - 1,2 A |
| Diagnostic test inlet open frequency reached | 145 Hz |
Installation
- Installation carton recycling : Le carton étant recyclable nous vous recommandons de l’éliminer selon les normes en vigueur. (Source : TURBO-SMART, TURBO-SMART TS, TURBO-SMART 2V, TURBO-SMART CUBE, p. 12)
- Installation dental unit contacts : Make sure that the contact is clean (no voltage should be present) (Source : TURBO-SMART, TURBO-SMART TS, TURBO-SMART 2V, TURBO-SMART CUBE, p. 15)
- Installation menu a5 wifi status : MENU WI-FI MENU COURT “A5” Vous pouvez vérifier l'état de votre connexion Wi-Fi. (Source : TURBO-SMART, TURBO-SMART TS, TURBO-SMART 2V, TURBO-SMART CUBE, p. 22)
- Installation menu drive status : Ce menu est accessible sans mot de passe et permet d’afficher des informations utiles concernant le fonctionnement. (Source : TURBO-SMART, TURBO-SMART TS, TURBO-SMART 2V, TURBO-SMART CUBE, p. 23)
- Installation menu fan switch off delay : Fan switch off delay (set 150 S) (Source : TURBO-SMART, TURBO-SMART TS, TURBO-SMART 2V, TURBO-SMART CUBE, p. 29)
- Installation menu generated code : Ceci est le code généré par Cattani S.p.A. Chaque appareil possède un code unique. (Source : TURBO-SMART, TURBO-SMART TS, TURBO-SMART 2V, TURBO-SMART CUBE, p. 27)
- Installation menu inverter calibration : Calibration inverter Réservé aux réglages en usine. (Source : TURBO-SMART, TURBO-SMART TS, TURBO-SMART 2V, TURBO-SMART CUBE, p. 30)
- Installation menu inverter status : Ce menu est accessible sans mot de passe et il permet de visualiser certains paramètres concernant l’état de fonctionnement de l'aspirateur. (Source : TURBO-SMART, TURBO-SMART TS, TURBO-SMART 2V, TURBO-SMART CUBE, p. 25)
- Installation menu minimum medium vacuum levels : Minimum vacuum level (only Micro Smart and Micro Smart Cube) (Source : TURBO-SMART, TURBO-SMART TS, TURBO-SMART 2V, TURBO-SMART CUBE, p. 28)
- Installation menu pump switch off delay : Pump switch off delay (max. 3600 t) (Source : TURBO SMART, TURBO SMART TS, TURBO SMART 2V, TURBO SMART CUBE, p. 29)
- Installation menu surgery vacuum level : Surgery vacuum level (only Micro Smart and Micro Smart Cube) (Source : TURBO-SMART, TURBO-SMART TS, TURBO-SMART 2V, TURBO-SMART CUBE, p. 29)
- Installation menu switch off delay : Switch off delay (max. 300 S) (Source : TURBO-SMART, TURBO-SMART TS, TURBO-SMART 2V, TURBO-SMART CUBE, p. 29)
User maintenance
- Consumable amalgam container : Replace with a new container (043132) (Source : Smart User Maintenance Guide (Cattani ESAM UK), p. 8)
- Consumable antifoaming tablets : Sanitising Antifoaming Tablets 040825 Stops foam build up, which can cause an interruption in the aspiration. 50 per box. (Source : Smart User Maintenance Guide (Cattani ESAM UK), p. 3)
- Consumable puli jet classic : Puli Jet Classic (4 x 5 litre bottle) 040715 Cleans, sanitises and deodorises dental suction units. 5 litres makes up to 250 litres of cleaner. (Source : Smart User Maintenance Guide (Cattani ESAM UK), p. 3)
- Consumable puli jet plus 2 0 range of action : Detergent disinfectant with bactericide, fungicide, virucide (HBV, HCV, HIV) and tuberculocide, proven effective in less than 15 minutes. Conforms to EN14476. Anti-scale. (Source : Smart User Maintenance Guide (Cattani ESAM UK), p. 3)
- Consumable pulse cleaner code : Pulse Cleaner 040720 Pulse Cleaner is a turbulence-maker. (Source : Smart User Maintenance Guide (Cattani ESAM UK), p. 3)
- Maintenance 12 months interval condition : EVERY 12 MONTH If required (routine maintenance not sufficient) (Source : Manutenzione straordinaria Turbo-Smart / Micro-Smart, p. 2)
- Maintenance amalgam 95 percent alarm : When the 95% level alarm is displayed, make arrangements to have the amalgam container replaced at an appropriate time (to do this the system will need to be turned off) (Source : Smart User Maintenance Guide (Cattani ESAM UK), p. 7)
- Maintenance daily disinfection : Jet plus 2.0 disinfectant with anti-scale agent (A) using the Pulse Cleaner (B) (Source : TURBO-SMART, TURBO-SMART TS, TURBO-SMART 2V, TURBO-SMART CUBE, p. 19)
- Maintenance dental unit filters : It is recommended that you clean the filters on the dental unit and replace the Antifoaming Tablets (if used) (Source : Smart User Maintenance Guide (Cattani ESAM UK), p. 6)
- Maintenance disinfect exterior : Désinfecter également l'extérieur de la machine. (Source : TURBO-SMART, TURBO-SMART TS, TURBO-SMART 2V, TURBO-SMART CUBE, p. 20)
- Maintenance every intervention temperatures and alarms : AT ANY INTERVENTION • Check maximum detected temperatures and all alarm signals* Intervene accordingly. (Source : Manutenzione straordinaria Turbo-Smart / Micro-Smart, p. 2)
- Maintenance occasional aspirator filter : Clean and disinfect the aspirator's filter (6) (Source : Manutenzione ordinaria Turbo-Smart / Micro-Smart, p. 2)