Technical data (Cattani)
| Rated voltage | 230 V |
| Supply phases | 1 |
| Maximum flow | 1600 l/min |
| Sound pressure level with canalized air | 70 dB(A) — Sound pressure level with canalized air 70 dB (A) |
| Sound pressure level measurement method | Sound pressure level detected according to ISO 3746-1979 (E). Parameters: r = 1.5, background noise 34 dB(A), Bruel & Kjaer type 2232 instrument. — Sound pressure level detected according to ISO 3746-1979 (E). Parameters: r= 1.5 - background noise: 34 dB(A) - Bruel & Kjaer type 2232 instrument. |
| Sound pressure model without box | 70 dB(A) |
| Width | 540 mm |
| Depth | 400 mm |
| Dimensions H | 650 mm |
| Protection against humidity | Common |
| Transport storage max humidity | Packages cannot be exposed to water and spraying and cannot withstand humidity above 70%. % |
| Diagnostic test inlet closed centrifuge separator current | 1,1 A |
| Diagnostic test inlet closed frequency reached | 50 - 70 Hz |
| Diagnostic test inlet closed head reached | 260 - 280 mbar |
| Diagnostic test inlet closed total current | 5,8 - 7 A |
| Diagnostic test inlet open centrifuge separator current | 1,1 A |
| Diagnostic test inlet open frequency reached | 85 - 110 Hz |
| Diagnostic test inlet open head reached | 55 - 75 mbar |
| Diagnostic test inlet open total current | 5,5 - 6,2 A |
| Display languages | Italian (0), English (1), French (2), Spanish (3), German (4) or Russian (5) |
| Drain pipe diameter | 18 mm |
| Use instructions | Continuous operation |
| Exhaust pipe diameter | 50 mm |
| Inverter frequency range | 70 Hz - 110 Hz Hz — Rotation speed 70 Hz 110 Hz |
| Operation suction motor | The fluid coming from the surgeries, enters the centrifugal separator (11) through the manifold (2) |
| Output power | 2 kW |
| Smartapp control | Turbo Smart, Turbo Smart TS, Turbo Smart 2V, Turbo Smart Cube, Micro Smart and Micro Smart Cube can be controlled by a professional using SmartApp. |
| Suction pipe diameter | 50 mm |
| Suction unit model | Uni Jet 75 2V motor |
| Transport storage stacking | The packages may be stacked only up to three rows having the same weight. — The packages may be stacked only up to three rows having the same weight |
Installation
- Installation carton recycling : Le carton étant recyclable nous vous recommandons de l’éliminer selon les normes en vigueur. (Source : TURBO-SMART, TURBO-SMART TS, TURBO-SMART 2V, TURBO-SMART CUBE, p. 12)
- Installation dental unit contacts : Make sure that the contact is clean (no voltage should be present) (Source : TURBO-SMART, TURBO-SMART TS, TURBO-SMART 2V, TURBO-SMART CUBE, p. 15)
- Installation menu a5 wifi status : MENU WI-FI MENU COURT “A5” Vous pouvez vérifier l'état de votre connexion Wi-Fi. (Source : TURBO-SMART, TURBO-SMART TS, TURBO-SMART 2V, TURBO-SMART CUBE, p. 22)
- Installation menu drive status : Ce menu est accessible sans mot de passe et permet d’afficher des informations utiles concernant le fonctionnement. (Source : TURBO-SMART, TURBO-SMART TS, TURBO-SMART 2V, TURBO-SMART CUBE, p. 23)
- Installation menu fan switch off delay : Fan switch off delay (set 150 S) (Source : TURBO-SMART, TURBO-SMART TS, TURBO-SMART 2V, TURBO-SMART CUBE, p. 29)
- Installation menu generated code : Ceci est le code généré par Cattani S.p.A. Chaque appareil possède un code unique. (Source : TURBO-SMART, TURBO-SMART TS, TURBO-SMART 2V, TURBO-SMART CUBE, p. 27)
- Installation menu inverter calibration : Calibration inverter Réservé aux réglages en usine. (Source : TURBO-SMART, TURBO-SMART TS, TURBO-SMART 2V, TURBO-SMART CUBE, p. 30)
- Installation menu inverter status : Ce menu est accessible sans mot de passe et il permet de visualiser certains paramètres concernant l’état de fonctionnement de l'aspirateur. (Source : TURBO-SMART, TURBO-SMART TS, TURBO-SMART 2V, TURBO-SMART CUBE, p. 25)
- Installation menu pump switch off delay : Pump switch off delay (max. 3600 t) (Source : TURBO SMART, TURBO SMART TS, TURBO SMART 2V, TURBO SMART CUBE, p. 29)
- Installation menu switch off delay : Switch off delay (max. 300 S) (Source : TURBO-SMART, TURBO-SMART TS, TURBO-SMART 2V, TURBO-SMART CUBE, p. 29)
- Installation menu system parameters setup : Il est possible de modifier le retard à l'extinction ainsi que d’autres paramètres techniques. (Source : TURBO-SMART, TURBO-SMART TS, TURBO-SMART 2V, TURBO-SMART CUBE, p. 24)
- Installation menu user parameters : Il est possible de modifier le niveau de vide, la langue, de lire le code généré et d’accéder au menu Activation. (Source : TURBO-SMART, TURBO-SMART TS, TURBO-SMART 2V, TURBO-SMART CUBE, p. 23)
User maintenance
- Consumable antifoaming tablets : Sanitising Antifoaming Tablets 040825 Stops foam build up, which can cause an interruption in the aspiration. 50 per box. (Source : Smart User Maintenance Guide (Cattani ESAM UK), p. 3)
- Consumable heat resistant tubing : 230212 Heat resistant tubing (precut) (Source : SMART_SYSTEMS_MAINTENANCE_SHEETS_DAILY&ANNUAL_UK&IRELAND.pdf (Cattani ESAM UK), p. 5)
- Consumable puli jet classic : Puli Jet Classic (4 x 5 litre bottle) 040715 Cleans, sanitises and deodorises dental suction units. 5 litres makes up to 250 litres of cleaner. (Source : Smart User Maintenance Guide (Cattani ESAM UK), p. 3)
- Consumable puli jet plus 2 0 range of action : Detergent disinfectant with bactericide, fungicide, virucide (HBV, HCV, HIV) and tuberculocide, proven effective in less than 15 minutes. Conforms to EN14476. Anti-scale. (Source : Smart User Maintenance Guide (Cattani ESAM UK), p. 3)
- Consumable pulse cleaner code : Pulse Cleaner 040720 Pulse Cleaner is a turbulence-maker. (Source : Smart User Maintenance Guide (Cattani ESAM UK), p. 3)
- Maintenance amalgam 95 percent alarm : When the 95% level alarm is displayed, make arrangements to have the amalgam container replaced at an appropriate time (to do this the system will need to be turned off) (Source : Smart User Maintenance Guide (Cattani ESAM UK), p. 7)
- Maintenance daily disinfection : Jet plus 2.0 disinfectant with anti-scale agent (A) using the Pulse Cleaner (B) (Source : TURBO-SMART, TURBO-SMART TS, TURBO-SMART 2V, TURBO-SMART CUBE, p. 19)
- Maintenance dental unit filters : It is recommended that you clean the filters on the dental unit and replace the Antifoaming Tablets (if used) (Source : Smart User Maintenance Guide (Cattani ESAM UK), p. 6)
- Maintenance disinfect exterior : Désinfecter également l'extérieur de la machine. (Source : TURBO-SMART, TURBO-SMART TS, TURBO-SMART 2V, TURBO-SMART CUBE, p. 20)
- Maintenance puli jet water volume : Fill the Pulse Cleaner Bucket with 2.5 litres of warm water (approx temp 50°c) (Source : Smart User Maintenance Guide (Cattani ESAM UK), p. 5)